Combined 1H MRI, PET and Multinuclear MRS hybrid imaging system

In this abstract we describe a novel method of combining three imaging modalities: 1H Magnetic Resonance Imaging (MRI), Positron Emission Tomography (PET), and multinuclear Magnetic Resonance Spectroscopy (MRS) in one system dedicated for true molecular imaging.

The addition of PET to MRI was introduced to provide functional/metabolic information about diseases or natural processes in the human body. High resolution and soft tissue contrast of 1H MRI morphological images is complemented by PET’s ability to depict metabolic processes through the use of biologically active radioactively labeled imaging agents, paving the way for molecular imaging.

The application of localized 1H MRS is implemented by using the imaging coil tuned to proton resonances. Localized, or MRI-guided, MR spectroscopy enables quantification of chemical composition of different tissue components in small volumes of less than 0.1 ml. Particularly in the brain, 1H MRS can for example measure accurately N-acetyl aspartate (NAA), lactate, glutamate, creatine, and choline to provide a view of the progression of neuro-degeneration.

However, the use of coil exclusively tuned to 1H significantly limits the “window” of observation by elimination of other biologically relevant nuclei like 13C, 14N, 17O, or 31P. Therefore we propose introducing a second resonator that will be tuned to these nuclei. We anticipate that MRS of other than 1H nuclei will provide more spectroscopic details and remove the ambiguity that exists in the 1H spectra from overlapping lines. The additional coil tuned to specific X-nuclei can be embedded within the existing 1H coil or can be removable, to be added whenever MRS on these nuclei is deemed necessary.

Author information

Authors and Affiliations

Authors

Rights and permissions

Open Access  This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made.

The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder.

To view a copy of this licence, visit https://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Hankiewicz, J.H., Celinski, Z., Smiley, K. et al. Combined 1H MRI, PET and Multinuclear MRS hybrid imaging system. EJNMMI Phys 1 (Suppl 1), A6 (2014). https://doi.org/10.1186/2197-7364-1-S1-A6

Download citation

  • Published:

  • DOI: https://doi.org/10.1186/2197-7364-1-S1-A6

Keywords