A robust MR-based attenuation map generation in short-TE MR images of the head employing hybrid spatial fuzzy C-means clustering and intensity inhomogeneity correction

Deriving an accurate attenuation correction map (μ-map) from magnetic resonance (MR) volumes has become an important problem in hybrid PET/MR imaging. Recently, short echo-time (STE) MR imaging technique incorporating fuzzy C-means (FCM) tissue classification and 2-point Dixon image acquisition has been introduced as a feasible approach for segmentation of the bone from air and soft tissue. However, this method imposes additional imaging and the performance of the standard FCM algorithm, suffering from the lack of spatial information, becomes impaired in the presence of inherent noise and intensity inhomogeneity. Here, we exploit a spatial fuzzy C-means (SFCM) segmentation algorithm in combination with a robust intensity inhomogeneity correction method on single STE-MR images, to differentiate various tissue classes.

MR images of five subjects were acquired on a clinical 1.5T MRI System, MAGNETOM Avanto, using a FLASH 3D pulse sequence with TE=1.1ms, TR=12ms, flip angle=18°, voxel size=1.2×1.2×2mm3.

The proposed segmentation approach consists of four main steps: (1) Intensity-inhomogeneity correction, to separate air and bone in the regions with high inhomogeneity, like the nasal areas; (2) applying SFCM to segment the images into four clusters including air, a part of soft tissue and bone, and two other soft tissue classes. Upon this step, the air cluster would be accurately separated; (3) employing shape factor analysis to remove the eyes with close signal intensity to that of bone; and (4) μ-map generation by downsampling and assigning attenuation coefficients to the corresponding segmented tissues.

Quantitative evaluation indicated sensitivity of over 95% for air and soft tissue and about 81% in the bony region, and specificity of over 95% for all tissue classes. The proposed STE-MR imaging in combination with the segmentation technique could be potentially exploited as an efficient approach to generate MR-based attenuation correction maps in clinical PET/MR applications, where bone and air coexist.

Author information

Authors and Affiliations

Authors

Rights and permissions

Open Access  This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made.

The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder.

To view a copy of this licence, visit https://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Aarabi, M.H., Kazerooni, A.F., Khateri, P. et al. A robust MR-based attenuation map generation in short-TE MR images of the head employing hybrid spatial fuzzy C-means clustering and intensity inhomogeneity correction. EJNMMI Phys 1 (Suppl 1), A49 (2014). https://doi.org/10.1186/2197-7364-1-S1-A49

Download citation

  • Published:

  • DOI: https://doi.org/10.1186/2197-7364-1-S1-A49

Keywords