Figure 3

From: Missing head and color banding in low-count SPECT reconstructions

Figure 3

Comparison of three transaxial Astonish reconstructions with scatter and attenuation correction. From left to right, Astonish without upscaling, 12× upscaling, and 100× upscaling of the projection data. The color scale is scaled with the same factor as the projection data. Identical slice position as in Figure 1. The images are quite similar, but the middle panel shows color banding, which is not present in the other panels. Upscaling more as in the right panel removes this color banding but introduces minor differences compared to the non-upscaled image in the left panel. The image in the left panel was scaled by the software before storing.