Skip to main content

Table 4 Sensitivity, specificity, accuracy, and area under curve of composite SUVR for each reconstruction

From: Phantom and clinical evaluation of the Bayesian penalised likelihood reconstruction algorithm Q.Clear without PSF correction in amyloid PET images

Reconstruction

Sensitivity

Specificity

Accuracy

Area under curve

Cut-off

OSEM

93.1

92.9

93.0

0.969

0.505

BPL with PSF

96.3

97.7

97.2

0.995

0.505

BPL without PSF

93.3

92.7

93.0

0.969

0.495