Skip to main content

Table 1 System specifications

From: Reduction of [68Ga]Ga-DOTA-TATE injected activity for digital PET/MR in comparison with analogue PET/CT

 

GE Signa PET/MR

Siemens Biograph mCT PET/CT

Scintilator

Lutetium‐yttrium oxyorthosilicate (LYSO)

Lutetium oxyorthosilicate (LSO)

Photodetector

SiPM

PMT

TOF (ps)

< 400

< 540

Axial view FOV (cm)

25

21.6

Trans-axial FOV(cm)

60

78

peak NECR (kcps @ kBq/ml)

216.8 @ 18.6

181.0 @ 25.2

Sensitivity at 10 cm offset (cps/kBq)

21

10

  1. PMT photomultiplier, ps picoseconds, FOV: field of view, NECR: noise equivalent count rate, cps: counts per second, kBq: kilobecquerel