Skip to main content

Table 2 Validation of our in-house NEMA spatial resolution tool

From: Validation of a discovery MI 4-ring model according to the NEMA NU 2-2018 standards: from Monte Carlo simulations to clinical-like reconstructions

Profile direction

GE console (mm)

In-house tool (mm)

Relative difference (%)

FWHM

FWTM

FWHM

FWTM

FWHM (%)

FWTM (%)

1 cm radial offset

 Radial

4.47

8.86

4.34

8.48

− 2.9

− 4.3

 Tangential

4.29

8.57

4.17

8.20

− 2.8

− 4.3

 Axial

4.63

10.23

4.49

10.07

− 2.9

− 2.1

10 cm radial offset

 Radial

5.58

10.65

5.66

11.30

1.5

6.1

 Tangential

4.53

9.76

4.40

9.77

− 2.8

0.2

 Axial

6.23

11.98

6.21

11.79

− 1.0

− 1.6

20 cm radial offset

 Radial

7.40

14.14

7.27

14.35

− 1.8

1.5

 Tangential

5.18

9.72

5.10

9.88

− 1.4

1.6

 Axial

6.27

12.24

6.29

12.45

0.4

1.8