Table 1 Technical characteristics of DIQ and DMI PET/CT systems

From: Insertion of synthetic lesions on patient data: a method for evaluating clinical performance differences between PET systems

 

DIQ

DMI

Scintillator crystals

BGO

LYSO

Crystal size (mm3)

6.3 × 6.3 × 30

3.95 × 5.3 × 25

Light amplification

Analogic (PMT)

Digital (SiPM)

Time-of-flight (TOF)

No

Yes

Configuration (number of ring)

5-rings

5-rings

Axial/transverse field of view (FOV) (mm)

250/700

250/700