Table 4 Parameters describing the system sensitivity as function of source collimator distance

From: Feasibility of 177Lu activity quantification using a small portable CZT-based gamma-camera

 

Energy window [keV]

\({c}_{0}\) [cps/MBq]

\({c}_{1}\) [cps/MBq]

\({c}_{2}\) [mm−1]

MEGP (gross)

55

6.8

5.7

0.0025

Large cylinder

113

− 1.2

18.4

0.0008

 

208

7.8

5.6

0.0079

MEGP (net, TEW)

55

1.6

4.9

0.0008

Large cylinder

113

− 12.1

25.1

0.0003

 

208

6.3

4.4

0.0067

LEHR (gross)

55

14.7

5.3

0.0143

Large cylinder

113

20.2

7.0

0.0107

 

208

12.6

16.6

0.0232

LEHR (net, TEW)

55

5.5

3.4

0.0027

Large cylinder

113

12.4

6.1

0.0036

 

208

10.1

12.8

0.0227

MEGP (gross)

55

5.5

6.6

0.0021

Small tube

113

9.4

8.0

0.0023

 

208

8.1

5.1

0.0101

MEGP (net, TEW)

55

5.7

0.6

0.0026

Small tube

113

4.7

8.6

0.0009

 

208

6.5

4.0

0.0090

LEHR (gross)

55

14.5

4.1

0.0198

Small tube

113

20.7

5.9

0.0157

 

208

12.1

11.0

0.0306

LEHR (net, TEW)

55

7.4

1.2

0.0082

Small tube

113

13.2

5.7

0.0046

 

208

9.8

8.5

0.0300

  1. Parameter values are given for MEGP and LEHR collimators, for gross count rates and net count rates after TEW scatter correction, and for the large cylindrical source and the small tube source