From: Feasibility of 177Lu activity quantification using a small portable CZT-based gamma-camera
Collimator | MEGP | LEHR | ||||||||
---|---|---|---|---|---|---|---|---|---|---|
Sphere volume/mL | 7.8 | 15.3 | 7.8 | 15.3 | ||||||
Camera position | a | a | b | a | b | a | a | b | a | b |
Reference depth/mm | 24 | 67 | 71 | 61 | 65 | 24 | 67 | 71 | 61 | 65 |
Activity deviation w TEW/% | − 4 | − 13 | − 15 | − 13 | − 17 | 7 | − 10 | − 14 | − 10 | 5 |
Activity deviation wo TEW/% | 3 | − 10 | − 13 | − 14 | − 17 | 25 | − 4 | − 8 | 0 | 9 |
Depth deviation w TEW/mm | 2 | 5 | 4 | 11 | 7 | 8 | 6 | 3 | 8 | 12 |
Depth deviation wo TEW/mm | 5 | 9 | 5 | 8 | 5 | 11 | 12 | 7 | 13 | 13 |