From: Quantitative 177Lu SPECT/CT imaging for personalized dosimetry using a ring-shaped CZT-based camera
Reconstruction | Recovery coefficient [%] (cold background) | Recovery coefficient [%] (hot background) | |||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|
10 mm | 13 mm | 17 mm | 22 mm | 28 mm | 37 mm | 10 mm | 13 mm | 17 mm | 22 mm | 28 mm | 37 mm | ||
GE-OSEM | 192i1s | 44 | 49 | 56 | 58 | 62 | 69 | 22 | 45 | 41 | 53 | 59 | 68 |
12i16s | 45 | 47 | 52 | 58 | 61 | 71 | 22 | 48 | 43 | 53 | 60 | 69 | |
Q.Clear | 192i1s | 33 | 46 | 53 | 55 | 58 | 67 | 17 | 30 | 29 | 41 | 50 | 64 |
12i16s | 44 | 46 | 52 | 57 | 61 | 70 | 22 | 44 | 41 | 52 | 59 | 68 | |
Q.ClearRDP | 192i1s | 22 | 39 | 46 | 51 | 55 | 65 | 15 | 24 | 24 | 35 | 46 | 60 |
12i16s | 28 | 39 | 44 | 51 | 56 | 67 | 17 | 29 | 28 | 41 | 52 | 63 |