Table 3 Spatial resolution and noise for GE-OSEM, Q.Clear and Q.ClearRDP with 192i1s and 12i16s

From: Quantitative 177Lu SPECT/CT imaging for personalized dosimetry using a ring-shaped CZT-based camera

Reconstruction

Spatial resolution (cold bgr) [mm]

Spatial resolution (hot bgr) [mm]

Noise

GE-OSEM

192i1s

7.8

10.8

0.29

12i16s

7.8

10.6

0.30

Q.Clear

192i1s

9.2

13.8

0.16

12i16s

8.0

10.8

0.27

Q.ClearRDP

192i1s

10.6

15.6

0.12

12i16s

10.4

13.6

0.15