Skip to main content

Table 2 Percentage error in the quantification of the activity in VOI_outside and of the activity concentration in VOI_inside for the cylinder phantom scan reconstructed with GE-OSEM, Q.Clear and Q.ClearRDP for different numbers of iterations and 1 subset

From: Quantitative 177Lu SPECT/CT imaging for personalized dosimetry using a ring-shaped CZT-based camera

Algorithm

Metric

12i1s

24i1s

48i1s

96i1s

192i1s

288i1s

576i1s

GE-OSEM

Error VOI_outside [%]

− 1.8

− 1.8

− 1.4

− 1.0

− 1.1

− 1.2

− 1.2

Error VOI_inside [%]

3.8

− 4.0

− 8.5

− 9.0

− 7.8

− 7.6

− 7.6

Q.Clear

Error VOI_outside [%]

− 1.8

− 1.8

− 1.6

− 1.4

− 1.2

− 1.2

− 0.6

Error VOI_inside [%]

4.8

− 1.7

− 6.8

− 8.8

− 8.8

− 8.7

− 7.9

Q.ClearRDP

Error VOI_outside [%]

− 1.9

− 1.8

− 1.8

− 1.6

− 1.4

− 1.5

− 0.8

Error VOI_inside [%]

4.8

− 1.6

− 6.8

− 8.9

− 9.1

− 9.1

− 8.3