Skip to main content

Table 1 Image calibration factors for GE-OSEM, Q.Clear and Q.ClearRDP for different numbers of iterations and 1 subset

From: Quantitative 177Lu SPECT/CT imaging for personalized dosimetry using a ring-shaped CZT-based camera

Algorithm

Image calibration factor [cps/MBq]

12i1s

24i1s

48i1s

96i1s

192i1s

288i1s

576i1s

GE-OSEM

89.5

93.2

93.8

94.0

93.9

93.9

93.8

Q.Clear

87.9

92.6

93.6

93.9

93.9

94.0

94.0

Q.ClearRDP

87.9

92.6

93.7

94.0

94.1

94.2

94.2