Table 3 SPECT/CT acquisition parameters

From: On the use of solid 133Ba sources as surrogate for liquid 131I in SPECT/CT calibration: a European multi-centre evaluation

Radionuclide

133Ba cylinders

131I cylinders

131I uniform Jaszczak

Collimator

High-energy

Number of energy windows

3

Photopeak energy (keV)

356.0 ± 26.7

(15% width)

364.5 ± 36.6

(20% width)

364.5 ± 36.6

(20% width)

Lower scatter energy (keV)

321.0 ± 8.0

(5% width)

317.1 ± 9.5

(6% width)

317.1 ± 9.5

(6% width)

High scatter energy (keV)

403.0 ± 20.2

(10% width)

411.9 ± 12.4

(6% width)

411.9 ± 12.4

(6% width)

Flood uniformity

As per clinical imaging protocol for 131I

Matrix

128

SPECT orbit

Body contour

Number of projections

120 (60 per detector)

Time/projection

30 s

30 s

60 s

Detector movement

Step-and-shoot

CT

Standard low-dose protocol