Table 1 Results from the survey among Dutch hospitals with a nuclear medicine department

From: Gamma camera-specific reference standards for radioactive iodine uptake measurements

Survey questions

Academic medical center #1

Academic medical center #2

Community hospital #1

Community hospital #2

Community hospital #3

Community hospital #4

Community hospital #5

Community hospital #6

1. Do you perform Iodine-131 therapy (e.g., for treatment of hyperthyroidism, multi-nodular goiter of toxic adenoma)?

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

2. Which radionuclide do you use as reference source?

Iodine-123

Iodine-131

Iodine-123

Iodine-123

Iodine-123

Iodine-123

Iodine-123

Iodine-123

3. What amount of activity (in MBq) do you use for the reference source?

11.1

0.5

11.1

11.1

18.5

37

18.5

11.1

4. Which formulation do you use for the reference source?

Capsule

Capsule

Capsule

Capsule

Capsule

Capsule

Capsule

Capsule

5. Do you use a phantom for measurement of the reference source? If so, what kind of phantom?

Neck phantom with a compartment for a capsule

Neck phantom with a compartment for a standard vial

Neck phantom with a compartment for a standard vial

No phantom

Neck phantom with a compartment for a standard vial

Neck phantom with a compartment for a capsule

Neck phantom with a compartment for a standard vial

Neck phantom with a compartment for a capsule

6. How do you prepare the reference source (before placement in the phantom)?

Direct placement of the capsule inside the phantom

Dissolvement of the capsule in (warm) water in a standard vial

Dissolvement of the capsule in (warm) water in a standard vial

Dissolvement of the capsule in (warm) water in a standard vial

Dissolvement of the capsule in (warm) water in a standard vial

Direct placement of the capsule inside the phantom

Dissolvement of the capsule in (warm) water in a standard vial

Direct placement of the capsule inside the phantom

7. How often is a new reference source prepared?

Daily

Daily

Weekly

Weekly

Prior to each patient

Daily

Weekly

Weekly

8. How often is the reference source measured?

Daily

Prior to each patient

Weekly

Weekly

Prior to each patient

Prior to each patient

Daily

Prior to each patient

9. What distance (in cm) between phantom and collimator head is used?

10

11.5

6.7

10

10

10

12

10

10. How long is the reference source measured?

1 min

2 min

100 kilocounts

5 min

1 min

5 min

10 min

1 min

11. Do you perform a background measurement?

No

Yes, in a separate background scan

No

Yes, in a separate background scan

Yes, in the neck phantom image

No

Yes, in a separate background scan

No

12. What method is used for determining the number of counts of the reference source?

Manual ROI

Total counts in the whole image

Automatic ROI

Square box ROI

Square box ROI

Automatic ROI

Automatic ROI

Automatic ROI

13. What type of equipment do you use for RAIU measurement?

Siemens Symbia T6

Siemens Symbia and Symbia Intevo

Siemens Symbia

Siemens Symbia S or Symbia T

Siemens Symbia and Symbia Intevo

Siemens Symbia T and Symbia Intevo

Siemens Symbia T2

Siemens Symbia T2

Survey questions

Community hospital #7

Community hospital #8

Community hospital #9

Community hospital #10

Community hospital #11

Community hospital #12

Community hospital #13

Community hospital #14

Community hospital #15

1. Do you perform Iodine-131 therapy (e.g., for treatment of hyperthyroidism, multi-nodular goiter of toxic adenoma)?

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

2. Which radionuclide do you use as reference source?

Iodine-131

Iodine-123

Iodine-123

Iodine-131

Iodine-123

Iodine-123

Iodine-123

Iodine-123

Iodine-123

3. What amount of activity (in MBq) do you use for the reference source?

10

18.5

18.5

0.5

3.7

10

11.1

3.7

18.5

4. Which formulation do you use for the reference source?

Syringe

Capsule

Capsule

Capsule

Capsule

Syringe

Capsule

Capsule

Capsule

5. Do you use a phantom for measurement of the reference source? If so, what kind of phantom?

No phantom

No phantom

Neck phantom with a compartment for a standard vial

Neck phantom with a compartment for a standard vial

Neck phantom with a compartment for a standard vial

No phantom

Neck phantom with a compartment for a capsule

No phantom

No phantom

6. How do you prepare the reference source (before placement in the phantom)?

Not applicable

Dissolvement of the capsule in (warm) water in a standard vial

Dissolvement of the capsule in (warm) water in a standard vial

Dissolvement of the capsule in hydrochloric acid

Direct placement of the capsule in a vial inside the phantom

Placement of syringe on shoulder of patient

Direct placement of the capsule inside the phantom

Not applicable

Not applicable

7. How often is a new reference source prepared?

One every two weeks

Daily

Weekly

Daily

Prior to each patient

Prior to each patient

Weekly

Prior to each patient

Daily

8. How often is the reference source measured?

Prior to each patient

Daily

Weekly

Prior to each patient

Prior to each patient

Prior to each patient

Weekly

Prior to each patient

Prior to each patient

9. What distance (in cm) between phantom and collimator head is used?

Similar distance to the patient scan procedure

10

10

25

20

As close to patient as possible

6

Not applicable

13

10. How long is the reference source measured?

1 min

5 min

1 min

5 min

5 min

2 min

1 min

Reference source is scanned together with the patient

1 min

11. Do you perform a background measurement?

Yes, in a separate background scan

No

Yes, in the neck phantom image

Yes, in the patient image

Yes, in a separate background scan

Yes, in the patient image

Yes, in the neck phantom image

Yes, in the patient image

No

12. What method is used for determining the number of counts of the reference source?

Automatic ROI

Automatic ROI

Square box ROI

Probe, Iodine-131 peak (330–418 keV)

Total counts in the whole image

Square box ROI

Automatic ROI

Square box ROI

Automatic ROI

13. What type of equipment do you use for RAIU measurement?

MED Nuklear-Medizintechnik upt2000

Siemens Symbia S

Siemens Symbia S and Symbia Intevo

Canberra Unispec

Siemens Intevo

Siemens Symbia T

GE NM/CT 870 DR

GE 640/670

Siemens Symbian Intevo