Table 3 Upper and lower limits of RCs and their ranges in EARL2 [26]

From: Variability in PET image quality and quantification measured with a permanently filled 68Ge-phantom: a multi-center study

 

Diameter of the hot sphere (mm)

 

10

13

17

22

28

37

RCmax

 Upper

0.88

1.22

1.38

1.32

1.26

1.29

 Lower

0.52

0.85

1.00

1.01

1.01

1.05

 Range (upper–lower)

0.36

0.37

0.38

0.31

0.25

0.24

RCmean

 Upper

0.61

0.86

0.97

0.99

0.97

1.00

 Lower

0.39

0.63

0.76

0.80

0.82

0.85

 Range (upper–lower)

0.22

0.23

0.21

0.19

0.15

0.15

RCpeak

 Upper

0.41

0.70

0.99

1.10

1.10

1.10

 Lower

0.27

0.45

0.75

0.90

0.90

0.90

 Range (upper–lower)

0.14

0.25

0.24

0.20

0.20

0.20