Skip to main content

Table 4 Experimental and simulated RC values from OSEM algorithm for each SBR and sphere size

From: Evaluation of a method based on synthetic data inserted into raw data prior to reconstruction for the assessment of PET scanners

 

OSEM

Experimental series

Simulated series

SBR 2:1

SBR 4:1

SBR 6:1

SBR 8:1

SBR 2:1

SBR 4:1

SBR 6:1

SBR 8:1

Sphere

Mean recovery coefficient: Average ± SD

13 mm

0.67 ± 0.04

0.65 ± 0.02

0.64 ± 0.01

0.60 ± 0.01

0.71 ± 0.03

0.66 ± 0.02

0.64 ± 0.01

0.63 ± 0.01

10 mm

0.65 ± 0.03

0.58 ± 0.02

0.57 ± 0.02

0.56 ± 0.01

0.69 ± 0.04

0.63 ± 0.02

0.61 ± 0.02

0.60 ± 0.01

8 mm

0.56 ± 0.04

0.45 ± 0.03

0.47 ± 0.01

0.45 ± 0.01

0.58 ± 0.04

0.53 ± 0.03

0.51 ± 0.03

0.50 ± 0.02

6 mm

0.43 ± 0.06

0.46 ± 0.01

0.40 ± 0.01

0.36 ± 0.03

0.49 ± 0.05

0.43 ± 0.03

0.40 ± 0.02

0.38 ± 0.02

5 mm

0.42 ± 0.08

0.32 ± 0.06

0.30 ± 0.02

0.29 ± 0.05

0.45 ± 0.06

0.35 ± 0.04

0.31 ± 0.02

0.30 ± 0.02

4 mm

0.42 ± 0.14

0.29 ± 0.03

0.32 ± 0.05

0.24 ± 0.05

0.40 ± 0.06

0.31 ± 0.06

0.26 ± 0.05

0.23 ± 0.04

Sphere

Maximum recovery coefficient: Average ± SD

13 mm

1.67 ± 0.29

1.22 ± 0.02

1.12 ± 0.02

0.99 ± 0.03

1.49 ± 0.08

1.26 ± 0.04

1.16 ± 0.04

1.11 ± 0.03

10 mm

1.46 ± 0.24

1.11 ± 0.06

1.05 ± 0.11

0.93 ± 0.06

1.49 ± 0.17

1.27 ± 0.13

1.16 ± 0.11

1.10 ± 0.09

8 mm

1.11 ± 0.08

0.83 ± 0.15

0.83 ± 0.03

0.87 ± 0.06

1.08 ± 0.23

0.92 ± 0.03

0.84 ± 0.04

0.75 ± 0.10

6 mm

0.70 ± 0.04

0.79 ± 0.13

0.70 ± 0.13

0.63 ± 0.03

0.98 ± 0.07

0.81 ± 0.03

0.74 ± 0.04

0.65 ± 0.11

5 mm

0.67 ± 0.23

0.48 ± 0.08

0.47 ± 0.05

0.42 ± 0.10

0.72 ± 0.06

0.55 ± 0.05

0.49 ± 0.06

0.44 ± 0.09

4 mm

0.55 ± 0.14

0.38 ± 0.07

0.41 ± 0.11

0.30 ± 0.08

0.62 ± 0.16

0.42 ± 0.08

0.34 ± 0.07

0.31 ± 0.05