Fig. 3From: Positron range in combination with point-spread-function correction: an evaluation of different implementations for [124I]-PET imagingComparison of the reconstructed (a) NEMA Image Quality phantom with OSEM (2 iterations and 12 subsets), OSEM+PRC simplified (2 iterations and 12 subsets), OSEM+PRC (8 iterations and 12 subsets), PSF (8 iterations and 12 subsets), PSF+PRC simplified (7 iterations and 12 subsets), and PSF+PRC (15 iterations and 12 subsets) and (b) Small-tumor phantom with OSEM (2 iterations and 12 subsets), OSEM+PRC simplified (2 iterations and 12 subsets), OSEM+PRC (9 iterations and 12 subsets), PSF (8 iterations and 12 subsets), PSF+PRC simplified (7 iterations and 12 subsets) and PSF+PRC (18 iterations and 12 subsets). For the NEMA IQ phantom, the highest differences (%) seen were for the smaller spheres 10, 13, and 17 mm and the edge of the phantomBack to article page