Table 2 Summary of spatial resolution of intraoperative gamma detectors used in the present study and detectors reported in the literature [27, 3032, 34, 35]. Zaburlini et al. [34] and Haigh et al. [32] included detector B in their evaluations and their results for this detector is shown separately

From: Evaluation of two intraoperative gamma detectors for assessment of 177Lu activity concentration in vivo

 

This study

Zamburlini et al. [34]

Haigh et al. [32]

Johnsrud et al. [35]

Benjegård et al. [31]

Britten [30]

Waddington et al. [27]

Radionuclide

177Lu

99mTc

99mTc/111In/125I

99mTc

111In

99mTc

111In

Number of detectors

2

6

5

1

3

3

2

 

FWHM (mm)

Depth (mm)

Detector A

Detector B

Detector B

All detectors

Detector B

All detectors

    

0

11

9.9

  

35/27/24

12–35/18–28/16–28

20–30a

5.2–8.7

  

5

        

11–16

 

10

27

20

     

8.3–23

13–20

 

20

41

30

     

11–35

19–30

12–43

30

  

35

20–60

    

26–41

 

40

68

49

        

50

        

38–56

23–82

60

93

73

        

80

113

91

        
  1. aTwo measurements were made for the same probe, one with and one without an extra collimator