Skip to main content

Table 1 Spatial resolution, given as full-width-at-half-maximum (FWHM) and full-width-at-tenth-maximum (FWTM), for the two detectors when the 177Lu line source was placed at different depths in the phantom of tissue-equivalent plastic. For some depths, FWTM was larger than 140 mm and could thus not be determined

From: Evaluation of two intraoperative gamma detectors for assessment of 177Lu activity concentration in vivo

 

FWHM (mm)

FWTM (mm)

Depth (mm)

Detector A

Detector B

Detector A

Detector B

0

11

9.9

39

18

10

27

20

88

46

20

41

30

124

77

40

68

49

>140

135

60

93

73

>140

>140

80

113

91

>140

>140