First results with SiPM tiles for TOF PET based on FBK RGB-HD technology

  • Alessandro Ferri1,
  • Fabio Acerbi1,
  • Peter Fischer2,
  • Alberto Gola1,
  • Giovanni Paternoster1,
  • Claudio Piemonte2,
  • Michael Ritzert2,
  • Torsten Solf3,
  • Illaria Sacco2 and
  • Nicola Zorzi1
EJNMMI Physics20152(Suppl 1):A86

DOI: 10.1186/2197-7364-2-S1-A86

Published: 18 May 2015

We present the first results of timing and energy resolution of two newly developed tiles based on FBK RGB-HD SiPMs. The first tile has dimensions of 32x32 mm2 and is composed of 8x8 SiPMs, with a regular pitch of 4 mm and a cell size of 25x25 μm2. Although manufactured with a standard bond wire technology, the tile achieves a fill factor at the tile level of 85%. We produced two versions: one with a single-ended and the other with differential readout. We tested the first prototypes with single-ended readout with a scintillator array, perfectly matching the tile pitch and composed of 8x8 LYSO crystals with dimensions of 4x4x22 mm3. First, we tested the tile using a single-channel setup, based on a fast, discrete amplifier, a digital oscilloscope and a PC, reading one SiPM at a time. At 20 °C, we measured an energy resolution of 10.7% FWHM. For the timing measurements we compared two conditions: when only one SiPM was biased and read, and when all the 64 SiPMs were biased but only one was read. At 20 °C, we measured a timing resolution of 200 ps FWHM in the first case, and 220 ps FWHM in the second case. Then, we tested the whole tile with a dedicated ASIC (PETA3), and measured the energy and timing resolution of two tiles in coincidence. The second tile is composed of 144 SiPMs, mounted on a water cooled, ceramic LTCC substrate. On the top side, it contains 12x12 SiPMs with a regular pitch of 2.5 mm. Also in this case, the SiPM technology is the RGB-HD with a cell size of 25x25 μm2. On the bottom side, four readout ASICs of the latest generation (PETA5) are flip-chip mounted. First results will be presented.

Authors’ Affiliations

(1)
Fondazione Bruno Kessler
(2)
Heidelberg University
(3)
Philips Healthcare

Copyright

© Ferri et al; licensee Springer. 2015

This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.