The development of MRI compatible detectors based on compact solid state photomultipliers has recently led to simultaneous fully integrated PET/MRI systems for human imaging. The PET acquisition design for MRI integration is known to have several additional constraints, including smaller space, electromagnetic compatibility issues and thermal management. The current work presents the PET acquisition architecture that has been developed for the TRIMAGE project, whose aim is to provide a cost effective, commercial grade trimodality PET/MRI/EEG scanner. The TRIMAGE PET component consists of 216 modules of 2.5 cm x 2.5 cm, arranged in 18 rectangular detectors of 5 cm x 15 cm, the latter in the axial direction, to form a full ring of 31 cm diameter. Each module consists of a staggered dual layer LYSO matrix read out by two arrays of 4 x 8 SiPMs and an ASIC. The detector board hosts a low-power low-end FPGA that performs pixel identification, energy calibration and handles the communication between the ASICs and the motherboard, which is located in proximity of the scanner. Data is streamed using high-density shielded cables and high-speed LVDS transmission to 9 low-end SoC FPGAs and from there to a central mainboard where coincidences and events statistics are processed. Coincidence data is finally transmitted to a host PC for image reconstruction. The proposed architecture and technological solutions will be presented and discussed.

Author information

Authors and Affiliations

Authors

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (https://creativecommons.org/licenses/by/4.0), which permits use, duplication, adaptation, distribution, and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Sportelli, G., Belcari, N., Bisogni, M.G. et al. A single-mode data acquisition architecture for PET/MRI. EJNMMI Phys 2 (Suppl 1), A19 (2015). https://doi.org/10.1186/2197-7364-2-S1-A19

Download citation

  • Published:

  • DOI: https://doi.org/10.1186/2197-7364-2-S1-A19

Keywords