Measurement of the point spread function of a pixelated detector array

  • Christian Ritzer1,
  • Patrick Hallen1,
  • David Schug1 and
  • Volkmar Schulz1
EJNMMI Physics20152(Suppl 1):A10

DOI: 10.1186/2197-7364-2-S1-A10

Published: 18 May 2015

In order to further understand the PET/MRI scanner of our group, we measured the point spread function of a preclinical scintillation crystal array with a pitch of 1 mm and a total size of 30 mm ~ 30 mm ~ 12 mm. It is coupled via a lightguide to a dSiPM from Philips Digital Photon Counting, used on the TEK-setup. Crystal identification is done with a centre of gravity algorithm and the whole data analysis is performed with the same processing software as for the PET insert, giving comparable results. The beam is created with a 22 NA-Point-Source and a lead collimator, with 0.5 mm bore diameter. The algorithm sorted 62 % of the coincidences into the correct crystal.

Authors’ Affiliations

(1)
Department of Physics of Molecular Imaging Systems, Institute for Experimental Molecular Imaging, RWTH Aachen University

Copyright

© Ritzer et al; licensee Springer. 2015

This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.