PET performance evaluation of a preclinical digital PET/MRI insert

In recent years, several groups have presented combined PET/MRI devices for clinical and preclinical research. Using new technologies like SiPMs, the performance of the systems improved over the years allowing a detailed exploration of the relatively new field of hybrid imaging modality. First presented at the NSS/MIC 2012, our group built a PET/MRI insert based on PDPC’s digital SiPMs (DPC). It is to date the only combined PET/MRI system using this technology which allows the performance evaluation of DPCs on system level. The Hyperion-IID PET/MR insert consists of 10 PET modules. Each PET module comprises 6 detector stacks whereas one of them consists of an interface board with local FPGA and a sensor tile. The latter employs a 4x4 array of sensor dies, each consisting of 2x2 DPCs. For photon detection, a 30x30 LYSO crystal array (1mm pitch, 12mm length) is glued using an intermediate 2mm light guide on top of each sensor tile. Measurements were performed using 22Na point sources (point like, activity = 1.5MBq) for different scenarios: the position dependence of the mentioned parameter was tested by moving a point source along the z axis. Different sensor configurations (trigger settings, validation thresholds, etc.) as well as temperature conditions were applied to characterize the influence of these parameters on the performance. We use an Anger algorithm and a Maximum Likelihood algorithm for crystal identification.

We measure 12.3% (FWHM) energy resolution and a CRT of about 470ps and 520ps (FWHM) for trigger schemes 2 and 3. No dependence on the axial position is observed and no difference between the measurement inside and outside the magnetic field occurs. We measure for both B0 field scenarios an effective sensitivity of about 3.1%.

Only minor effects due to the operation inside the B0 field can be observed. More results and details about this study will be presented at the conference.

Author information

Authors and Affiliations

Authors

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (https://creativecommons.org/licenses/by/4.0), which permits use, duplication, adaptation, distribution, and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Schug, D., Lerche, C., Dueppenbecker, P. et al. PET performance evaluation of a preclinical digital PET/MRI insert. EJNMMI Phys 1 (Suppl 1), A3 (2014). https://doi.org/10.1186/2197-7364-1-S1-A3

Download citation

  • Published:

  • DOI: https://doi.org/10.1186/2197-7364-1-S1-A3

Keywords