Hybrid PET/MRI insert: B0 field optimization by applying active and passive shimming on PET detector level

Combining PET and MRI into a hybrid device is challenging since both systems might influence each other. A typical interference problem of such a combined device is the distortion of the MRI’s B0 field distribution due to the material brought inside the MRI’s FOV which is in particular challenging for small-bore PET-systems. High field homogeneity is needed for a good MRI acquisition in general as well as in certain applications. Typically, active shimming using dedicated coils is applied to improve the field homogeneity. However, these techniques are limited especially for localized distortion profiles with higher-order characteristics caused by PET/MRI inserts. As a consequence, we are exploring the potential application of shimming on PET detector level (for the Hyperion-IID PET/MRI insert), meaning that the distortion profile caused by PET modules is compensated using additional magnetic materials (passive shimming) and DC coils (active shimming). To explore the technique, B0 field measurements have been performed using a whole-body phantom in combination with the MRI body coil. An FFE sequence was used to measure distortion maps of DC loops and small magnetic objects (capacitors, ferrites). These distortion maps served as input for a software framework which has been written to perform the field optimization. The implementation was verified by measurements and fits were performed to extract characteristic parameters of the tested objects. Finally, the implemented software framework was used to homogenize a measured distortion map produced by a single PET module by superimposing distortion corrections from additional simulated materials. The resulting superimposed distortion map showed a significantly improved B0 field map quality (reduced spectral width and improved homogeneity). The simulated susceptibility distribution will be applied on PET module level and tested in experiments. Results and details about this study will be presented at the conference.

Author information

Authors and Affiliations

Authors

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (https://creativecommons.org/licenses/by/4.0), which permits use, duplication, adaptation, distribution, and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Wehner, J., Weissler, B. & Schulz, V. Hybrid PET/MRI insert: B0 field optimization by applying active and passive shimming on PET detector level. EJNMMI Phys 1 (Suppl 1), A15 (2014). https://doi.org/10.1186/2197-7364-1-S1-A15

Download citation

  • Published:

  • DOI: https://doi.org/10.1186/2197-7364-1-S1-A15

Keywords